Rheon eyeWatch Pen (Standard / Single-Use)
Available

Rheon eyeWatch Pen (Standard / Single-Use)

Reference: C4, sC1

The eyeWatch Pen is a reusable external control device used to non-invasively adjust the eyeWatch glaucoma drainage implant after surgery. By rotating the pen around the circumference of the implanted eyeWatch, the surgeon activates the internal magnetic disc to compress or decompress the drainage tube, thereby modulating aqueous humor outflow and intraocular pressure. The eyeWatch Pen enables precise, office-based IOP titration without additional surgical intervention. Its ergonomic design allows single-handed operation for efficient postoperative patient management. Sterilizable and designed for repeated clinical use across multiple patients.

Technical sheet

Reference: C4 (eyeWatch Pen, office version); sC1 (eyeWatch Pen single-use)

Type: external device for reading and adjusting the eyeWatch implant

Functions: reading of implant position (compass); non-invasive adjustment (external magnet)

Positions: 0 (fully open) to 6 (fully closed)

Magnet: magnetic field about 150 times the Earth's magnetic field

Office version (C4): supplied non-sterile, post-operative adjustments

Single-use version (sC1): sterile (X-ray), intra-operative reading and adjustment, supplied with the eyeWatch implant

Manufacturer: Rheon Medical SA (Switzerland)