Opério® / SteriStay® — Laminar airflow generators
The Opério® and SteriStay® generators protect the surgical site and instruments from airborne contamination, one of the causes of surgical site infections. Room air passes through a HEPA H14 filter and a sterile laminar fabric: the ultra-clean air is directed without turbulence over the risk zone and pushes contaminated air away. The sterile zone is shown by a light beam, and single-use sterile drapes are checked by a barcode reader. Three versions: Opério® Mobile, a mobile unit with a removable instrument tray; Opério® Ceiling, a ceiling-mounted unit on an articulated arm that keeps the floor clear; SteriStay®, a mobile, height-adjustable instrument table with filtered airflow. They work independently of the room's ventilation system.
Technical sheet
Opério® Mobile: 230 V, 290 W (standby 23 W); 45 x 60 x 170 cm; 42 kg; optional tray 45 x 60 or 40 x 50 cm, adjustable from 80 to 120 cm, 5 kg max; working field depth 120 cm
Opério® Ceiling: 230 V, 160 W (standby 23 W); length 60 cm, reach 180 to 200 cm, width 25 cm, height 90 cm; 35 kg; working field depth 100 cm
SteriStay®: 230 V, 160 W (standby 23 W); 170 x 60 x 170 cm; 60 kg; table 133 x 60 cm (S version 173 x 60 cm), adjustable from 80 to 120 cm, 50 kg max; working field depth 133 cm
HEPA class H14 filter: 99.995% MPPS; lifetime 1 year or 2,000 hours; nylon mesh pre-filter
Airflow: 0.4 to 0.5 m/s, 400 m³/h; clean zone < 5 CFU/m³
Sterile zone pointer: class 3R laser (5 mW, green 520 nm); class 2 laser barcode reader (1 mW, red 650 nm)
Conditions: 0 to 30 °C, 5 to 80% humidity non-condensing, 500 to 1,000 hPa
Compliance: CE, Medical Devices Directive 93/42/EEC