NIDEK yellow-wavelength (577 nm) laser photocoagulator, slit-lamp mounted. YLC-500 single spot, YLC-500 VIXI multispot.

Technical sheet

Reference: YLC-500

Manufacturer: NIDEK

Type: 577 nm yellow photocoagulation laser (YLC-500 single spot / YLC-500 VIXI multispot)

Source: diode-pumped solid-state laser

Wavelength: 577 nm

Emission mode: continuous, pulses 0.01 to 0.2 s

Output power at cornea: 50 to 1700 mW (single spot); 50 to 1500 mW (VIXI multispot)

Exposure time: 0.01 to 3.00 s (single spot); 0.01 to 0.05 s (multispot)

Repeat interval: 0.05 to 1.00 s

Cooling: air

Aiming beam: 635 nm laser diode (red), continuous, 0.2 to 0.4 mW

Spot size: 50 to 1000 µm (single spot); 100 to 500 µm in multispot, 50 to 500 µm single spot (VIXI)

Multispot patterns (VIXI): square and equidistant 2×2 to 5×5, rectangle, triple arc, triple curve, macular grid, triangle, circle, circular arcs, curve, line

Delivery systems: on NIDEK SL-1800 slit lamp, detachable (NIDEK SL-2000 / SL-1800 / SL-1600, ZEISS SL-130; HAAG-STREIT 900 BQ for multispot), HEINE OMEGA 500 binocular indirect ophthalmoscope (single spot)

Power supply: 100–240 V AC ±10%, 50–60 Hz

Power consumption: less than 250 VA

Console dimensions: 237 × 318 × 90 mm (single spot); 276 × 318 × 90 mm (multispot)

Console weight: 5.6 kg (single spot); 7.15 kg (multispot)

Footprint with SL-1800 slit lamp and table: 760 × 450 × 1300 to 1500 mm, about 45 kg